Nitrogen gas discharge system

This illustration depicts the HAKKO FM-2026.
What is the temperature of heated nitrogen gas?
- In case of HAKKO FM-2026 (Station:HAKKO FX-951)
- Test condition:
- Measure at a point 2mm from the center of tip end with ø0.1CA sensor
- Tip shape
- T13-D24
- Temperature
- 300℃ and 350℃

Setting sample
-Soldering station HAKKO FX-951
-N2 generator HAKKO FX-780:
small-capacity type
-N2 flow meter HAKKO FX-791
Setting |
Flow rate of |
Temperature of |
---|---|---|
300 |
0.5 |
210 |
1.0 |
250 |
|
1.5 |
260 |
|
2.0 |
263 |
|
2.5 |
260 |
|
350 |
0.5 |
230 |
1.0 |
240 |
|
1.5 |
300 |
|
2.0 |
320 |
|
2.5 |
305 |
- In case of HAKKO FX-8302 (Station:HAKKO FX-838)
- Test condition:
- Measure at a point 2mm from the center of tip end with ø0.1CA sensor
- Tip shape
- T20-D24
- Temperature
- 300ºC and 350ºC

Setting sample
-Soldering station HAKKO FX-838
-N2 generator HAKKO FX-780:
small-capacity type
-N2 flow meter HAKKO FX-791
Setting temperature |
Flow rate of |
Temperature of |
---|---|---|
300 |
0.5 |
225 |
1.0 |
235 |
|
1.5 |
240 |
|
2.0 |
175 |
|
2.5 |
160 |
|
350 |
0.5 |
275 |
1.0 |
300 |
|
1.5 |
300 |
|
2.0 |
225 |
|
2.5 |
200 |